TM 11-6625-2398-15-3
Step
Test equipment
Equipment under
Test
Performance
No.
control setting
test control settings
procedure
standard
POWER: OFF
zero set the OUTPUT VOLTS
(down)
meter.
OUTPUT VOLTS:
zero set
d. SG-340/G
POWER: ON
adjust the OUTPUT LEVEL control
Attenuator: 0db
knob for an OUTPUT VOLTS
OUTPUT VOLTS:
meter indication at the SET LEVEL
SET LEVEL
(+4DBM) mark.
Note: Check that OUTPUT
VOLTS meter indication .is exactly
on the SET LEVEL mark with the
attenuator set to 0dB (outer scale).
Readjust the OUTPUT LEVEL con-
trol knob if required.
2
SG-340/G
None
Adjust the MOD. LEVEL control
Function Switch:
for a reading of 50 on the PER-
1000
CENT MODULATION meter. Ad-
PERCENT
just the standing wave ratio indi-
MODULATION: 50
cator VERNIER GAIN control and
MOD. UP NORMAL
RANGE switch for an exact meter
DOWN:
reading (e.g. .5 on the EXPANDED
NORMAL
DB meter scale). Record the meter
AN UPM-108A
reading.
METER SCALE:
EXPAND
RANGE
As required for meter reading
3
Controls remain as at end of step 2.
a. Disconnect the hybrid attenuator
under test and connect the variable
attenuator, as shown by broken
b. Adjust the variable attenuator
Insertion loss on the
ATTENUATOR DB control for the
ATTENUATOR DB
same standing wave indicator meter
control of the variable
reading recorded in step 2 above.
attenuator must
indicate 6.51.5.
Note. Correct the
hybrid attenuator
marking for 1090MHz, if
incorrect.
4
SG-340/G
None
a. Reconnect the hybrid attenuator,
MEGACYCLES:
solid lines figure 18-3.
1030
Balance of controls
Insertion loss on the
remain as at end of
MEGACYCLES dial to 1030 with
ATTENUATOR DB
step 2.
control Of the vari-
peat steps 2 and 3.
able attenuator must
indicate. 6.51.5.
Note. Correct the
hybrid attenuator
for 1030MHz.
if incorrect.
8-12