TM 11-6625-2398-15-3
CHAPTER 9
DEPOT OVERHAUL STANDARDS
and will be capable of conforming to their respective
9-1. Applicability of Depot Overhaul Standards
repair standard or depot overhaul standard.
Test Facilities Set AN/TPM-24 (V) 3 must be tested
thoroughly after rebuild or repair to ensure that it meets
Qty.
Applicable
Test equipment
rqd.
literate
adequate performance standards for return to stock and
reissue. Use the test described in this chapter and in the
1
TM 11-2050
Coaxial Slotted Line IM-92/U
1
TM 11-5109
It is mandatory that equipment to be reissued, or returned
Indicator, Standing Wave
1
TM 11-6625-335-12
Ratio AN/UPM-108
1
TM 11-6625-335-12
to stock for reissue, meet all of these performance
standards.
PSM-6A, and AN/PSM-6B
1
TM 11-6625-475-0
9-2. Applicable References
a. Repair Standards. Applicable procedures of the
(A)/G (Model 612A UHF
depot performing this and its general standards for
1
repaired equipment form a part of the requirements for
Variable Attenuator Alfred
testing this equipment,
Electronics Model E101
1
b. Technical Publications. The following Technical
Crystal Detector HP-423A
1
Manuals are applicable to this equipment.
Low Pass Filter HP-360C
1
Adapter (type N male to type
BNC female) UG-606./U
2
Adapter (type N male to type
Equipment and subject
Publication
BNC female) UG-201A/U
2
DS, GS, and Depot Maintenance
TM 11-5895-687-35-1
Adapter (type BNC female to
Manual Interrogator Set AN/
BNC female) UG-914/U
2
TPX-50
50 ohm termination. Micro-
DS, GS, and Depot Maintenance
TM 11-5895-687-35-2
lab/FXR TA-6MN
2
Manual Signal, Processor CP-39g/
9-4. General Test Requirements
TPX-50
Always allow at least 30 minutes for the equipment under
DS, GS, and Depot Maintenance
TM 11-5895-687-35-3
test and the test equipment to reach stabilized
Manual Synchronizer, Electrical
temperatures. The test equipment and the materials
8N-421/TPX-50
required to perform the following test are listed in
DS, GS, and Depot Maintenance
TM 11-5896-687-35-4
Radio RT-903/TPX-5o
Check the continuity of all cables, extender boards, and
c. Modification Work Orders. Perform all applicable
front panel adapter.
Modification Work Orders (MWO's) pertaining to this
9-6. Insulation Resistance
equipment before making test specified. DA Pam 310-7
lists all available MWO's
Measure the insulation resistance of all cables (the
9-3. Test Facilities Required
resistance between the conductor and the insulation
The following equipments, or suitable equivalents of
material). A minimum reading of 100 megohms is
known accuracy, will be employed in determining
required.
compliance with the requirements of this specific standard
9-1