C5, TM 11-6625-168345
cable
assembly
connector contact and the connector assembly
(19) CX-12067/U
frame, and between the tip of the telephone
(AUDIO).
plug and the housing of the plug. Leakage
should be greater than 10 megohms. High-pot
connector contact A and the tip of the tele-
at 500 volts ac for 1 minute.
phone plug, and between contact B and the
stem pf the telephone plug. Use the RX1 range
(20) CX-10441A/PPS-5 cable assembly
( block 2100). Follow the test procedures in
should be less than 0.1 ohm.
(3) (a) and (b) above.
(b) Perform a leakage and high-pot
(21) CX-10444A/PPS-5 cable assembly
test between connector contacts A and B, and
(block 400). Follow the test procedures in
between the tip and the stem of the telephone
(16) (a) and (b) above.
plug. Perform the same test between each
8.1-14.1